Resolution is typically defined in military standards as
maximal spatial frequency of a standard USAF 1951 target that can be revolved by
an observer at a certain illuminance level and the target contrast
Spot defects in form of dark spots are the most common type
of blemishes present in images generated by I2 tubes, and also in I2 systems.
The defects are defined as the opaque or dark spots which exceed contrast of 30
percent of their surrounding area. It is typically required that the dark spots
shall not exceed the size and quantities specified in special tables presented
in MIL standards.
Attention: It is recommended to measure dark spots during
testing image intensifier tubes (see ITS-I test station) because then it is
possible to carry out objective measurement with higher accuracy. During testing
image intensifier systems only subjective evaluation is possible.
System gain (luminance gain) is a parameter defined as the
ratio of the output screen luminance to the input illuminance at the
target plane. This parameter gives information how much brighter is the tubes
screen seen by the ocular in comparison to the average illumination at the
target plane.
Minimum Resolvable Contrast function is a function of a
minimum contrast difference between the bars of the standard target and the
background required to resolve the image of the bars by an observer versus
spatial frequency of the target at different levels of target luminance. The
USAF 1951 target is used as the standard target during MRC measurement.
In other words MRC is at a resolution measured at different
illumination levels for different contrast levels; and the resolution is a point
at the MRC characteristic.
Because it is much more difficult to measure MRC than the
resolution, therefore resolution is more frequently measured than MRC.
Minimal resolvable contrast (MRC) function is considered as
the most important characteristic of visible imaging systems because having
known MRC function it is possible to calculate the ranges of detection,
recognition and identification of the targets of interest.
Distortion is a parameter that gives information about
distortion of image of different geometrical figures. There are many definitions
of distortion in literature. In case of distortion of I2 systems we recommend to
use definition of so the called gross distortion presented in MIL standards that
define parameters of I2 tubes. In this way we can defined distortion as a
phenomenon that causes the image of a straight line to curve.
Attention: It is recommended to measure dark spots during
testing image intensifier tubes (see ITS-I test station) because then it is
possible to carry out objective measurement with higher accuracy.
Magnification is a parameters defined as a ratio of the
angular size of the image of the target seen by the observer using the I2 system
(via the ocular) to the angular size of the true target seen using only a naked
eye.
Field Of View (FOV) is a parameter defined as an angular size
of the area of the observed scenery that can be seen using the I2 system. FOV is
determined by the useful size of the photocathode of the image intensifier tube
and the focal length of the input optics of the I2 system.
Collimation is an angle between optical axis of dual channel
goggles. When this angle is too high then the observer can have problems to use
both eyes for stereoscopic observation.